MP Filtri LPA2 Particle Counter

A unique high accuracy, fully portable product. For users of hydraulic, lubrication and transmission systems. The LPA2 is a highly accurate, portable laser particle analyser that counts and sizes particles of solid contaminants in fluid power systems, o
The experience of designers and users of hydraulic and lubrication systems is that 75% of system failures are as a direct result of contamination. Knowing the cleanliness level of the fluid is the basis for contamination control.
The LPA2 gives accuracte results of the amounts and sizes of contaminants - on the spot. LPA2 is calibrated with ISO Medium Test Dust (MTD) based on ISO 11171:1999 calibration standard. The new MTD has a certified distribution standard verified by NIST (National Institute of Standard and Technology), USA. The LPA2 is designed to meet the new ISO 4406 cleanliness classification code which is a 3-par code, 4µm(c), 6µm(c) and 14µm(c). The LPA2 also provides results in the NAS 1638 code.
- Calibration: Each unit is individually calibrated with ISO Medium Test Dust (MTD) as based on ISO 11171:1999.
- Analysis range: ISO 8 to ISO 24, ISO 4406 Code. (NAS 1638 Code - 2 to 12)
- Report/print format: ISO and NAS codes, with individual particle counts as a built-in option
- Printer: 36 column alpha/numeric impact printer
- LPA2 sample volume: 15 ml. (normal) 30 ml. (dynamic) 24 ml. (bottle sampler) 15ml. continuous
- Operation: Max. system working pressure - 400 bar. Min. working pressure - 2 bar.
- Viscosity range: to 400 centistokes
- Operating temperature: + 5 to + 80°C
- Fluid compatibility: Mineral oil & petroleum based fluids, and Skydrol® (consult MP FILTRI for other fluids)
- Typical test time: 2 mins.
- Power: Internal rechargeable battery (mains charger) or external 12/24 volt DC power supply.
- Data storage: 500 test
- Computer interface: RS 232 communication port
- Hose connections: Minimess fittings microbore hose 1.5 metres long waste fluid hose (5m -10m lengths available)
- Dimensions: Height 210 mm. Depth 260 mm. Width 430 mm. Weight: 7.6 kilos
- Optional Product: In line course screen unit minimess fitting. 500 micron st.steel cleanable mesh 400 bar filter pressure
- Technology Automatic optical particle analyser
- Laser package Twin laser and twin optical diode detectors
- LCD display (back lit)
- Sensitivity >4,6,14,21,25,38,50,68,µm(c), micron range to revised ISO 4406 Standard
- Accuracy/repeatability Better than 3% typical
- Includes Moisture & Sensor for online measurement of RH